Medical Policy
Subject: Implantable Left Atrial Hemodynamic Monitor
Document #: SURG.00128Publish Date: 09/27/2023
Status: ReviewedLast Review Date: 08/10/2023
Description/Scope

This document addresses implantable left atrial hemodynamic (LAH) monitors. These devices monitor left atrial pressure (LAP) with the objective of identifying pressure changes in ambulatory individuals with heart failure (HF) to potentially enable earlier intervention and prevention of clinical deterioration. The monitoring system can be used as a stand-alone device or in combination with an implantable cardioverter defibrillator (ICD) or cardiac resynchronization therapy defibrillator (CRT-D).

Note: For information regarding other technologies for cardiac disease, see:

Position Statement

Investigational and Not Medically Necessary:

Left atrial hemodynamic monitoring utilizing implantable device (for example, HeartPOD™ System, Promote® LAP System, and V-LAP™ System) is considered investigational and not medically necessary for all indications.

Rationale

The Hemodynamically Guided Home Self-Therapy in Severe Heart Failure Patients (HOMEOSTASIS) trial was a multicenter, prospective trial approved by the U.S. Food and Drug Administration (FDA) under an Investigational Devices Exemption (IDE) (Ritzema, 2007). This was the first-in-human trial of the HeartPOD System (Abbott Laboratories, Abbott Park, IL, formerly by St. Jude Medical/Savacor, Inc). The purpose was to evaluate the device’s safety, reliability, and functionality. Individuals with a history of New York Heart Association (NYHA) functional class III to IV HF and at least one hospital admission, emergency department or clinic visit for acute decompensated HF requiring a parenteral diuretic, vasodilator, or positive inotrope during the previous 12 months were eligible to participate. After device implantation, participants had clinic visits at 2, 6, and 12 weeks for a clinical assessment, data retrieval, and noninvasive device calibration. Concordance between pulmonary capillary wedge pressure (PCWP) and direct LAP from the device was assessed at week 12. A total of 8 individuals participated, and none experienced procedural complications. At week 12, 87% of HeartPOD LAP measurements were within ± 5 mm Hg of PCWP, and there were no device-related complications or deaths, unplanned clinic visits or hospital admissions for HF. The investigators concluded that HeartPOD could be safely implanted and accurately measures LAP in the short-term. They also stated that additional research is needed to determine whether direct LAP measurement can result in optimized HF treatment and thereby, improve clinical outcomes.

In 2010, Ritzema and colleagues reported additional results from the HOMEOSTASIS trial. This report included a total of 40 subjects who were implanted with the HeartPOD System and had a median follow-up of 25 months (ranging from 3-38 months). Primary safety endpoint measures were met at 6 weeks with all participants free of major adverse cardiac or neurological events; the 3-year survival rate for participants without decompensation was 61%, with a decrease in episodes reported after the initial 3 months. During pressure guided therapy, the reported mean daily LAP fell from 17.6 mm Hg to 14.8 mm Hg during the initial 3 months. The authors concluded that:

The small study size, lack of a randomized design with a concurrent control group, and observer bias from lack of blinding limit the ability to reach definitive conclusions about the safety and clinical effectiveness of this heart failure management strategy.

The 2022 American Heart Association (AHA)/American College of Cardiology (ACC)/Heart Failure Society of America (HFSA) guideline for the management of heart failure does not address use of implantable LAP devices (Heidenreich, 2022).

The Heart Failure Society of America Scientific Statements Committee published a white paper consensus statement on remote monitoring of patients with HF (Dickinson, 2018). It included an assessment of the LAPTOP-HF trial, which was a prospective, multicenter, randomized, unblinded trial of individuals who have been diagnosed with NYHA class III HF (Maurer, 2015). Participants were randomized to LAP-guided HF therapy (treatment group) or HF usual standard of care (control group). The primary endpoints assessed safety and effectiveness. The safety assessment included freedom from procedure/device-related major adverse cardiovascular and neurological events (MACNE) at 12 months. The effectiveness assessment included a composite endpoint of heart failure hospitalizations and complications of HF therapy over the entire study period. Preliminary results were presented at the 2016 Annual Scientific Meeting of the Heart Failure Society. The investigators reported that the planned enrollment was 730 individuals however, at 486 individuals, the trial was stopped by the Data and Safety Monitoring Board due to an excess of complications from the implantation procedure. In the absence of peer-reviewed data from a completed trial, the authors of this white paper concluded that:

However, after enrollment was terminated, the LAPTOP-HF trial’s steering committee recommended that all randomized individuals who were implanted successfully should be followed through at least 12 months. At the 2016 Annual Scientific Meeting of the Heart Failure Society, the preliminary results reported by Abraham and colleagues (2016) showed that freedom from MACNE at 12 months was 90.6% in the LAP-guided HF therapy group, with a lower confidence interval (CI) of 86.7% (the prespecified level was 80%). The annual heart failure hospital rate for the LAP-guided HF therapy group was 0.40 vs. 0.68 in the control group, which represented a relative risk reduction of 41%; p=0.005. The investigators concluded that ambulatory LAP-guided HF therapy was safe and associated with a 41% reduction in HF hospital admissions, and that this data provides insight into the potential benefits of hemodynamic monitoring. Additional randomized trials are needed to assess the safety and efficacy of this technology.

Perl and colleagues (2022) conducted the V-LAP Left Atrium Monitoring systEm for Patients With Chronic sysTOlic & Diastolic Congestive heart Failure (VECTOR-HF; NCT03775161) trial, the first-in-human trial of the V-LAP system. In this prospective, multicenter, open-label, single-arm trial, eligible participants were older than 18 years of age, diagnosed with chronic HF NYHA functional class III and had a history of at least one hospital admission due to worsening heart failure within the past year or elevated brain natriuretic peptide (BNP) > 300 pg/mL or N-terminal pro b-type natriuretic peptide (NT-proBNP) > 1500 pg/mL. Primary outcomes included successful implantation of the device, the ability to conduct initial pressure measurements, and device safety defined as freedom from major adverse cardiovascular and neurologic events. Secondary outcomes included accurate pressure measurements, transmission of information up to 3 months after implantation, device concordance with PCWP at 3 months, admissions due to heart failure, and changes in NYHA Class, 6-minute walk test, Kansas City Cardiomyopathy Questionnaire (KCCQ) results, and NT-proBNP levels at 6 months. A total of 24 individuals (83% were male) with a mean age of 67.4 ± 9.7 years underwent implantation with the V-LAP-System. All were successfully implanted with the device, and no device-related complications (defined as invasive treatment, device explant or death) or sensor failure occurred. Concordance between the V-LAP System and PCWP measurements showed a mean difference of -2.05 ± 3.33 mm Hg (Lin concordance correlation coefficient = 0.850, 95% CI, 0.676 to 0.934). At 6 months post implantation, 8 of 20 individuals had an improvement in NYHA Class (40%, 95% CI, 16.4% to 63.5%). There was no change in 6-minute walk test distance, KCCQ scores, NT-proBNP levels, or number of hospitalizations due to heart failure. The investigators stated that these initial results demonstrate that ambulatory hemodynamic monitoring with the V-LAP system is safe and feasible. They also stated that additional well-designed randomized trials are still needed.

Restivo and colleagues (2022) reported long-term outcomes of a subset of the VECTOR-HF study participants. The purpose was to describe the largest (n=5) and longest (median follow-up 18 months) single-center experience with the V-LAP device. Individuals who underwent V-LAP System implantation had an improvement in their 6-minute walk test distance (352.5 ± 86.2 meters at baseline to 441.2 ± 125.2 meters at last follow-up) and KCCQ overall scores 63.82 ± 16.36 vs. 81.92 ± 9.63 and KCCQ clinical score 68.47 ± 19.48 vs. 83.70 ± 15.58. The investigators concluded that their findings were promising however, additional research to confirm the device’s reliability as well as its clinical benefit is needed. 

Background/Overview

Description of Relevant Disease

According to the Centers for Disease Control and Prevention, approximately 6.2 million Americans are currently diagnosed with HF, and more than 960,000 new cases are diagnosed each year (CDC, 2020). Approximately 50% of individuals with HF die within 5 years of diagnosis. As a result of HF, the weakened heart muscle causes inadequate filling of the left ventricle, as well as a backflow of blood into the left atrium, both resulting in decreased cardiac output and increased symptoms for the afflicted individual. Symptoms can include shortness of breath, fatigue, swelling in the ankles, feet, legs, abdomen and veins in the neck. Currently there is no cure for HF; medical therapy includes a combination of diuretics, digoxin, angiotensin-converting enzyme (ACE) inhibitors, angiotensin receptor blockers (ARB), beta-blockers, and aldosterone antagonists. Some individuals may remain symptomatic, despite medical therapy. Ongoing studies evaluate other treatment options to assist physicians in the medical management of individuals with severe heart failure.

HeartPOD System:

This device is indicated for individuals with ischemic or non-ischemic cardiomyopathy with systolic or diastolic dysfunction for at least 6 months or HF classified by NYHA Class III. The HeartPOD system is a stand-alone device for use in individuals not requiring ICD or CRT-D therapy, or who already received ICD or CRT-D therapy. The system monitors LAP with a permanently implantable sensor used in ambulatory individuals with heart failure (HF). These implanted intracardiac sensors allow the individual to directly monitor left atrial pressure, the intracardiac electrogram, and core body temperature. The implant's readings are communicated with a hand-held computer called a PAM. The information is used to adjust medications on a dose-by-dose basis according to the physician's prescriptive instructions. The HeartPOD System is not available for commercial use in the U.S.

The Promote LAP System:

The Promote LAP System (St. Jude Medical, Inc., St. Paul, MN) is indicated for individuals with ischemic or non-ischemic cardiomyopathy and Class III HF. The Promote LAP system is a combination device for individuals who require ICD or CRT-D therapy in addition to LAP monitoring. This Promote LAP System is not available for commercial use in the U.S. Enrollment for an FDA IDE study was completed in 2016; study results have not been published.

V-LAP System:

The V-LAP System (Vectorious Medical Technologies, Tel Aviv, Israel) is an implanted device that is placed percutaneously across the inter-atrial septum and measures LAP. Individuals with HF may be monitored remotely through bidirectional communication with an external unit. Data from the device are transmitted to a Patient Advisor Module (PAM) that provides the individual with specific instructions on changes that should be made to their HF therapy. The changes are based on the individual’s hemodynamic measurements and physician’s directions. The goal of the device is to shift from crisis management to health maintenance by detecting an impending HF exacerbation (i.e., before the onset of symptoms) allowing for alterations in HF medications and thereby, avoiding complications. The V-LAP System is limited to investigational use in the U.S.

Definitions

Cardiomyopathy: A disease in which the heart muscle becomes inflamed and doesn't work as well as it should; there are three main types of cardiomyopathy:

Congestive Heart failure (CHF), also referred to as Heart Failure (HF): A condition in which the heart no longer adequately functions as a pump. As blood flow out of the heart slows, blood returning to the heart through the veins backs up, causing congestion in the lungs and other organs.

Investigational Device Exemption (IDE): Allows the investigational device to be used in a clinical study in order to collect safety and effectiveness data required to support a Premarket Approval (PMA) application or a Premarket Notification [510(k)] submission to FDA.

Ischemic dilated cardiomyopathy (IDCM): Left ventricular systolic dysfunction (or disease of the heart muscle) associated with at least 75 percent narrowing of at least one of the three major coronary arteries (marked stenosis) or a documented history of myocardial infarction.

New York Heart Association (NYHA) definitions: The NYHA classification of heart failure is a 4-tier system that categorizes based on subjective impression of the degree of functional compromise. The four NYHA functional classes are as follows:

Coding

The following codes for treatments and procedures applicable to this document are included below for informational purposes. Inclusion or exclusion of a procedure, diagnosis or device code(s) does not constitute or imply member coverage or provider reimbursement policy. Please refer to the member's contract benefits in effect at the time of service to determine coverage or non-coverage of these services as it applies to an individual member.

When services are Investigational and Not Medically Necessary:
When the code describes a procedure indicated in the Position Statement section as investigational and not medically necessary.

CPT

 

93799

Unlisted cardiovascular service or procedure [when specified as insertion of left atrial hemodynamic monitor]

 

 

ICD-10 Procedure

 

02H700Z-02H740Z

Insertion of pressure sensor monitoring device into left atrium [by approach; includes codes 02H700Z, 02H730Z, 02H740Z]

 

For the following codes when specified as left atrial monitoring

0JH600Z-0JH630Z

Insertion of hemodynamic monitoring device into chest subcutaneous tissue and fascia [by approach; includes codes 0JH600Z, 0JH630Z]

0JH800Z-0JH830Z

Insertion of hemodynamic monitoring device into abdomen subcutaneous tissue and fascia [by approach; includes codes 0JH800Z, 0JH830Z]

 

 

ICD-10 Diagnosis

 

 

All diagnoses

References

Peer Reviewed Publications:

  1. Abraham WT, Adamson PB, Costanzo MR, et al. Hemodynamic monitoring in advanced heart failure: Results from the LAPTOP-HF trial. J Card Fail. 2016; 22(11):940.
  2. Maurer MS, Adamson PB, Costanzo MR, et al. Rationale and design of the left atrial pressure monitoring to optimize heart failure therapy study (LAPTOP-HF). J Card Fail. 2015; 21(6):479-488.
  3. Perl L, Meerkin D, D'amario D, et al. The V-LAP System for remote left atrial pressure monitoring of patients with heart failure: Remote left atrial pressure monitoring. J Card Fail. 2022; 28(6):963-972.
  4. Restivo A, D'Amario D, Paglianiti DA, et al. A 3-year single center experience with left atrial pressure remote monitoring: The long and winding road. Front Cardiovasc Med. 2022; 9:899656.
  5. Ritzema J, Melton IC, Richards AM, et al. Direct left atrial pressure monitoring in ambulatory heart failure patients: initial experience with a new permanent implantable device. Circulation. 2007; 116(25):2952-2959.
  6. Ritzema J, Troughton R, Melton I, et al.; Hemodynamically Guided Home Self-Therapy in Severe Heart Failure Patients (HOMEOSTASIS) Study Group. Physician-directed patient self-management of left atrial pressure in advanced chronic heart failure. Circulation. 2010; 121(9):1086-1095.
  7. Troughton RW, Ritzema J, Eigler NL, et al. Direct left atrial pressure monitoring in severe heart failure: long-term sensor performance. J Cardiovasc Transl Res. 2011; 4(1):3-13.

Government Agency, Medical Society, and Other Authoritative Publications:

  1. Dickinson MG, Allen LA, Albert NA, et al. Remote monitoring of patients with heart failure: a White Paper From the Heart Failure Society of America Scientific Statements Committee. J Card Fail. 2018; 24(10):682-694.
  2. Heidenreich PA, Bozkurt B, Aguilar D, et al. 2022 AHA/ACC/HFSA Guideline for the management of heart failure: a report of the American College of Cardiology/American Heart Association Joint Committee on Clinical Practice Guidelines. Circulation. 2022. 145:e895-e1032.
  3. U.S. Food and Drug Administration. The Center for Devices and Radiological Health (CDRH). Investigational Devices Exemption (IDE). Available at: http://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/HowtoMarketYourDevice/InvestigationalDeviceExemptionIDE/default.htm. Accessed on June28, 2023.
Websites for Additional Information
  1. American Heart Association. Available at: http://www.americanheart.org. Accessed on June 28, 2023.
  2. Centers for Disease Control and Prevention. Heart failure. Last updated September 8, 2020. Available at: Heart Failure | cdc.gov. Accessed on June 28, 2023.
Index

HeartPOD System
Left atrial hemodynamic (LAH) monitor
Left atrial pressure (LAP) monitoring
Promote LAP System
V-LAP System

The use of specific product names is illustrative only. It is not intended to be a recommendation of one product over another, and is not intended to represent a complete listing of all products available.

Document History

Status

Date

Action

Reviewed

08/10/2023

Medical Policy & Technology Assessment Committee (MPTAC) review. Updated the Rationale, Background, References, Websites, and Index sections.

Reviewed

08/11/2022

MPTAC review. Updated Rationale, References and Websites sections.

Reviewed

08/12/2021

MPTAC review. Updated Rationale, Background, References and Websites sections.

Revised

08/13/2020

MPTAC review. Clarified INV and NMN statement, updating examples to include V-LAP™ System. Updated Rationale, References and Websites sections.

Reviewed

08/22/2019

MPTAC review. Updated Rationale, Background, References and Websites sections.

Reviewed

09/13/2018

MPTAC review. Updated Description, Rationale, References and Websites sections.

Reviewed

11/02/2017

MPTAC review. The document header wording updated from “Current Effective Date” to “Publish Date.” Updated Rationale, Background, References, and Websites sections. Updated Coding section with 01/01/2018 CPT changes; removed 0293T, 0294T deleted 12/31/2017.

Reviewed

11/03/2016

MPTAC review. Revised title: Implantable Left Atrial Hemodynamic Monitor. Updated Description, Rationale, Background, Definitions and Reference sections.

Reviewed

11/05/2015

MPTAC review. Updated Description, Rationale, References and Websites sections. Removed ICD-9 codes from Coding section.

Reviewed

11/13/2014

MPTAC review. Updated Background and References.

Reviewed

11/14/2013

MPTAC review. No change to Position Statement. Updated Rationale, Background, References and Websites.

Reviewed

11/08/2012

MPTAC review. Updated References and Websites.

New

11/17/2011

MPTAC review. Initial document development.


This Medical Policy provides assistance in understanding Healthy Blue’s standard Medicaid benefit plan. When evaluating coverage for a specific member benefit, reference to federal and state law, as well as contractual requirements may be necessary, since these may differ from our standard benefit plan. In the event of a conflict with standard plan benefits, federal, state and/or contractual requirements will govern. Before using this policy, please check all federal, state and/or contractual requirements applicable to the specific benefit plan coverage. Healthy Blue reserves the right to modify its Policies and Guidelines as necessary and in accordance with legal and contractual requirements. This Medical Policy is provided for informational purposes. It does not constitute medical advice. Healthy Blue may also use tools and criteria developed by third parties, to assist us in administering health benefits. Healthy Blue’s Policies and Guidelines are intended to be used  in accordance with the independent professional medical judgment of a qualified health care provider and do not constitute the practice of medicine or medical advice.
 
No part of this publication may be reproduced, stored in a retrieval system or transmitted, in any form or by any means, electronic, mechanical, photocopying, or otherwise, without permission from the health plan.  
 
© CPT Only – American Medical Association